• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550250-23
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Angina (1710); Stenosis (2263)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.The reported patient effects of angina and stenosis are listed in the xience sierra, everolimus eluting coronary stent system (eecss), electronic instructions for use as known patient effects of coronary stenting procedures.A conclusive cause for the reported angina and stenosis, and the relationship to the product, if any, cannot be determined.The subsequent treatment(s) of unexpected medical intervention (additional therapy/non-surgical treatment) appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling; therefore, no product-related corrective action will be implemented in this case.
 
Event or Problem Description
Patient id: (b)(6).It was reported that a 2.5x23 mm xience sierra stent was implanted in the mid left anterior descending (lad) coronary artery on (b)(6) 2020.The patient experienced angina on (b)(6) 2021.On (b)(6) 2022, during a follow up, 75% stenosis was noted in the proximal lad; however, no in-stent restenosis was noted in the xience sierra stent in the mid lad.On (b)(6) 2022, elective intervention was performed to treat the proximal lad and a 3.0x23 mm xience skypoint stent was implanted.The patient was discharged from the hospital the following day.According to the physician¿s opinion, the event is not related to the study as it is a treatment for de novo lesion occurring in the natural history; however, it is unknown if the restenosis was +/- 5 mm from the target lesion.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P031)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15789269
Report Number2024168-2022-11541
Device Sequence Number5472268
Product Code NIQ
UDI-Device Identifier08717648227110
UDI-Public08717648227110
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1550250-23
Device Catalogue Number1550250-23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2022
Initial Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight79 KG
-
-