| Catalog Number |
MB3555 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c device subsided after the patient went on an amusement park ride three months post-operatively.The patient is currently being treated non-operatively for new left shoulder and arm pain.However, a future revision to remove the mobi-c and convert it to an acdf has not been ruled out.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c device subsided at c4/5 after the patient went on an amusement park ride three months post-operatively.The patient was initially treated non-operatively for new left shoulder and arm pain.However, a revision surgery was later performed to remove the mobi-c devices at c3/4 and c4/5 and replace them with competitive product after the device at c3/4 migrated.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d4: catalog, lot, and udi numbers.Additional information in d4: expiration date, and h6: health effect.Reference report 3004788213-2023-00013.Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: product was not returned and no photos were provided.Product information is not known.A device evaluation, complaint history search and recall search are unable to be performed.Root cause: root cause was unable to be determined.This event could possibly be attributed to riding a roller coaster 3 months post-operatively, as mentioned in the complaint description.Dhr review: dhr review unable to be performed as lot number is not known.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c device subsided after the patient went on an amusement park ride three months post-operatively.The patient is currently being treated non-operatively for new left shoulder and arm pain.However, a future revision to remove the mobi-c and convert it to an acdf has not been ruled out.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: x-rays were provided which demonstrate the subsidence.The complaint is confirmed.Root cause: root cause was unable to be determined.This event could possibly be attributed to riding a roller coaster 3 months post-operatively, as mentioned in the complaint description.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c device subsided at c4/5 after the patient went on an amusement park ride three months post-operatively.The patient was initially treated non-operatively for new left shoulder and arm pain.However, a revision surgery was later performed to remove the mobi-c devices at c3/4 and c4/5 and replace them with competitive product after the device at c3/4 migrated.This is report one of two for this event.
|
| |
|
Search Alerts/Recalls
|