| Model Number |
1588610122 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.".
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: always getting foggy.[update - moisture could not be wiped away.Replacement camera used to complete procedure.] the failure(s) identified in the investigation is consistent with the complaint record.Root cause: leak at the strain relief attributed to the sterilization method used at the account.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Search Alerts/Recalls
|