Voluntary mdr report: mw5112877.An end user reported an issue with a mini stick max 5f x 10cm stiff.018 ni/tu echo 2.75" pg.The following was reported: "patient was brought to the angiography suite and placed in supine position.Initially transdural approach was attempted, and after radial artery was identified using ultrasound guidance and the overlying skin was infiltrated with 2cc of 1% lidocaine, next artery was accessed with a micro-puncture needle under direct ultrasound guidance and a micro-wire was advanced through the needle, however the micro-wire appeared to get stuck, and not advancing forward nor backwards.As the micro-wire was pulled back gently with the needle, it was noticed that the tip of the microwire broke off and got retained.Subsequent right distal forearm x-ray showed micro-wire position in the soft tissue, however interventionalist could not confirm if it was intravascular.Patient hand was monitored for several minutes and there was no evidence of ischemia or swelling.This unintended complication was communicated with the patient in detail, was made aware of the small part of wire that remained and was shown the pictures at the end of the procedure.The patient did not complain of any pain or distress and verbalized understanding.The case continued and femoral access obtained.
|
|
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of guidewire tip detached inside patient cannot be confirmed.No guidewire sample was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.Scar004647 was sent to guidewire supplier/manufacturer, heraeus medical for dhr review of supplier lot numbers.Heraeus confirmed no internal non-conformance reports were found regarding batch in question and personnel were properly trained.This event cannot be confirmed to be a manufacturing issue.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint. the review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: direction for use {dfu} is provided with this mini stick coaxial microintroducer.The failure mode of device fracture and embolization is cautioned as potential adverse event.No sample was returned for evaluation so it cannot be determined if device was used in a manner inconsistent with labeling.Adverse events guidewire shearing, fracture or embolization operational instructions: prior to insertion, flush all components with saline or heparinized/saline.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire product analysis: initial condition: n/a - no product returned to angiodynamics for evaluation.External evaluation: scar004647 has been initiated due to the reported complaint.The scar004647 was sent to the supplier, heraeus medical for dhr review of supplier lots {9009989284 and 9009989379}.See scar004647 for details.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|