| Model Number |
39-SP-0700 |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a procedure was on (b)(6) 2022, utilizing the reform pedicle screw system.Upon attempted insertion of one of the polyaxial screws the tip of the t20, polyaxial driver, reform degen (39-sp-0700) broke.The broken tip was left in the head of the screw, implanted in the patient.The rod was assembled over the screw with no issue.The doctor indicated that he has to use excessive force while inserting the screw.There was no patient injury or delay to the procedure reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Outcomes attributed to adverse event: other serious (important medical events).This is being reported as a serious injury (retained foreign object) due to the fractured tip of the driver being left in the head of the polyaxial screw implanted in the patient.It is important to note that once the rod implant is seated and locked into the tulip of the polyaxial screw, the fractured tip is prevented from dislodging as long as the rod remains firmly locked in place.Initial reporter occupation: other; distributor.Review of device history records is not possible without lot identification.Two-year complaint history review did not identify a trend for reports of this nature for the reported part number.Without the opportunity to evaluate the complaint product, no conclusions can be drawn.Initial information provided indicates that the product is available for evaluation but has yet to be returned to the manufacturer.Should the product be received, a follow-up medwatch report will be filed upon completion of evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that a procedure was on (b)(6) 2022, utilizing the reform pedicle screw system.Upon attempted insertion of one of the polyaxial screws the tip of the t20, polyaxial driver, reform degen (b)(6) broke.The broken tip was left in the head of the screw, implanted in the patient.The rod was assembled over the screw with no issue.The doctor indicated that he has to use excessive force while inserting the screw.There was no patient injury or delay to the procedure reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 device evaluation - the tip of the driver is fractured with the fracture plane primarily normal to the driver's longitudinal axis.The fracture surface was examined with the aid of a 5x loop.The surface includes multiple fracture planes.It is believed that the fracture may have been initiated during prior use based upon the surface appearance and was likely due to high torque application which was noted by the user.Review of device history records found thirty (30) pieces of lot 33611mm released for distribution on (b)(6) 2015 with no deviation or anomalies.Two-year complaint history review did not reveal a trend for reports of this nature for this part number.No corrective actions are recommended at this time.
|
| |
|