• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5MM HEALIX PEEK 3-SUTURE ANCHOR W/ORTHOCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5MM HEALIX PEEK 3-SUTURE ANCHOR W/ORTHOCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 222207
Medical Device Problem Codes Break (1069); Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by a healthcare professional in china that during an unknown surgery on (b)(6) 2022, it was observed that the suture on the 5.5mm healix peek 3-suture anchor w/orthocord device was cracked and broken off.Another like device was used to complete the surgery.There were no adverse consequences to the patient nor surgical delay reported.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device has not been returned, therefore unavailable for a physical evaluation.However, a photo was provided.Upon visual inspection of the photo, it appeared that suture was broken and frayed; therefore, this complaint can be confirmed.A manufacturing record evaluation was performed for the finished device lot number: 9l11158, and no nonconformances were identified.The photos do not provide enough evidence to determine root cause.Without physical evaluation of the devices reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause for the frayed suture can be attributed to procedures variables such as: snaps or other instrument to pull the suture without wrapping the suture around the snaps creates a stress point on the suture, resulting in breaking it, however, this cannot be conclusively determined.As per ifu, is indicated to apply tension (normal force) on suture lengths.Excessive force may overload the anchor or suture.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection of the device, it was observed a little piece of the suture was breakage and frayed on different parts.Therefore, this complaint can be confirmed.The anchor was not returned along with the device.Manufacturing record evaluation was performed for the finished device lot number and no nonconformances were identified.No additional information was provided about details of the procedure; therefore, a definitive root cause could not be determined.The possible root cause for the frayed suture can be attributed to procedures variables such as: snaps or other instrument to pull the suture without wrapping the suture around the snaps creates a stress point on the suture, resulting in breaking it, however, this cannot be conclusively determined.As per the instructions for use, it is indicated to apply tension (normal force) on suture lengths.Excessive force may overload the anchor or suture.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5MM HEALIX PEEK 3-SUTURE ANCHOR W/ORTHOCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15793983
Report Number1221934-2022-03548
Device Sequence Number11515720
Product Code MAI
UDI-Device Identifier10886705007554
UDI-Public10886705007554
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number222207
Device Catalogue Number222207
Device Lot Number9L11158
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2022
Supplement Date Received by Manufacturer11/20/2022
11/22/2022
01/13/2023
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date11/21/2022
11/23/2022
01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-