Model Number 150003
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the device manometer was not reading what it was set.No patient injury was reported.
 
Additional Manufacturer Narrative
Device evaluation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Operator of device is unknown, no information provided to date.
 
Additional Manufacturer Narrative
H10: device evaluation: one ventilator device was returned to manufacturing for investigation.Visual inspection and functional testing were performed.Visual inspection of the returned device confirmed that the tamper seal was intact upon receipt, the device was received with a bowed front panel on the left-hand side of the manometer.The relief pressure knob max end stop is 60.The diaphragm valve was missing.No error codes or event history log (ehl) are available for this device.After performing manometer checks using relief pressure settings, the customer reported issue was confirmed as the manometer needle does not match settings on the relief pressure knob.The root cause of this event is attributed to customer impact whereby the front panel was bent, the relief pressure end cap was out of alignment and the manometer tubing was found to be kinked internally.To correct this issue, the front panel and end cap alignment were re-shaped, and the manometer tubing was replaced.The previous service history has been reviewed and deemed unrelated to the current customer reported problem.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS BABYPAC
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15793985
Report Number3012307300-2022-27117
Device Sequence Number3010222
Product Code CBK
UDI-Device Identifier10610586045578
UDI-Public10610586045578
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K970158
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number150003
Device Catalogue Number150003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer05/23/2023
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date06/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-