Model Number TFGT-A
Medical Device Problem Code Intravalvular regurgitation (4068)
Health Effect - Clinical Code Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 10/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The implant date was estimated as 1 feb 2022; information was reported that the valve was implanted in (b)(6) 2022 on an unknown date.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that in (b)(6) 2019, a 19mm trifecta gt valve was successfully implanted during an aortic valve replacement.There was mild aortic regurgitation observed on the device, which was confirmed by echo aortography.No thrombus was observed on the device.A transcatheter aortic valve implantation, or valve-in-valve procedure, was scheduled for (b)(6) 2022.The patient was reported as stable.
 
Additional Manufacturer Narrative
An event of mild aortic regurgitation was reported.A more comprehensive assessment, including histopathological examination of the valve tissue could not be performed as the device remains implanted and was not returned for analysis.The device history record could not be reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications since no lot or serial number was provided.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Subsequent to the previously filed report, additional information was received: it was reported that the regurgitation was confirmed to be observed in october 2022.The valve-in-valve procedure proceeded on (b)(6) 2022 and was successful in implanting a 23mm non-abbott transcatheter valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA¿ GT VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL LTDA. REG#3001883144
1301rua profvieira demendonça
bairro engenho nogueira 31.31 0-26
BR   31.310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15794041
Report Number2135147-2022-02052
Device Sequence Number4546539
Product Code LWR
UDI-Device Identifier05415067018205
UDI-Public05415067018205
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P100029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFGT-A
Device Catalogue NumberTFGT-19A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/18/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date01/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-