• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558350
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant was unable to report the lot number; therefore, the manufacture date and expiration date are unknown.(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophageal dilatation performed on (b)(6) 2022.During the procedure, balloon inflation was attempted however, it could not be inflated.The device was removed from the patient and it was noticed that the balloon had ruptured.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophageal dilatation performed on (b)(6) 2022.During the procedure, balloon inflation was attempted however, it could not be inflated.The device was removed from the patient and it was noticed that the balloon had ruptured.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to report the lot number; therefore, the manufacture date and expiration date are unknown.Block h6: device code a0402 captures the reportable event of balloon burst.Block h10: investigation result: the returned cre fixed wire dilatation balloon was analyzed and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the distal section of the balloon.Microscopic inspection found the balloon had a pinhole located approximately 98mm from the tip.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon burst was not confirmed.The pinhole problem found could have been interpreted by the customer as the reported event of balloon burst.The pinhole found in the balloon is likely to have occurred due to factors encountered during the procedure: excess pressure, interaction with other devices, or patient anatomy.It is possible that factors encountered during the procedure, such as the interaction with any surface during the procedure could create friction on the balloon and cause a pinhole.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device was unable to be performed as the lot number is unknown.However, a ship history review was performed to identify the most probable lot numbers, and a manufacturing review of the most probable lot numbers did not identify any anomalies or deviations that could have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15794094
Report Number3005099803-2022-06623
Device Sequence Number4791264
Product Code KNQ
UDI-Device Identifier08714729202004
UDI-Public08714729202004
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K971320
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00558350
Device Catalogue Number5835
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer11/16/2022
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-