MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO¿ MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO¿ MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MZ9277
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an unspecified number of bd maxzero¿ multi-fuse extension set with needleless connectors had flow issues while injecting radioactive tracers, as well as during the flushing process.The following information was provided by the initial reporter: "the maxzero extension tubing creates back splash when injecting radioactive tracers.This causes significant contamination to he patient and unnecessary exposure to the caregiver staff and nursing teams on the floor.In addition to the contamination, the tracer, even when flushed, collects in the maxzero blue hub which reduces the image quality of our cardiac stress patients and potentially exposes nursing teams on the floor.".
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Additional Manufacturer Narrative
Investigation summary: no product or photo was returned by the customer.It was reported by the customer that the maxzero extension tubing creates back splash when injecting radioactive tracers could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed on model mz9277 because a lot number was not provided by the customer.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that an unspecified number of bd maxzero¿ multi-fuse extension set with needleless connectors had flow issues while injecting radioactive tracers, as well as during the flushing process.The following information was provided by the initial reporter: "the maxzero extension tubing creates back splash when injecting radioactive tracers.This causes significant contamination to he patient and unnecessary exposure to the caregiver staff and nursing teams on the floor.In addition to the contamination, the tracer, even when flushed, collects in the maxzero blue hub which reduces the image quality of our cardiac stress patients and potentially exposes nursing teams on the floor.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO¿ MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15795126
Report Number9616066-2022-01730
Device Sequence Number14650327
Product Code FPA
UDI-Device Identifier10885403240799
UDI-Public(01)10885403240799
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMZ9277
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-