STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1.5MM X 4CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035421540
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure of venous sinus emboli of dural-avf, the coil (subject device) was advanced into the microcatheter, but there was strong resistance about 10 cm before the tip of the catheter, so it was retrieved.While retrieving the coil (subject device) , it got prematurely detached at the rhv(rotating hemostasis valve).The detached coil (subject device) was safely removed from the patient¿s anatomy by removing the rhv.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual inspection of the subject coil device revealed the main coil was found to be detached/separated.Additionally, the core delivery wire was found to be kinked/ bent.The coil introducer sheath was not returned.The functional testing could not be performed due to the condition of the returned device.An assignable cause of procedural factors will be assigned to the reported and as analysed ' coil in catheter friction', as the issues are associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited the reported 'main coil prematurely detached/separated during use' was confirmed during the analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.As the main coil was found to be prematurely detached/separated and the coil delivery wire was found to be kinked/bent.Therefore, an assignable cause of procedural factors will be assigned to the reported and as analysed ¿ main coil prematurely detached/separated during use¿, as the issues are associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.A probable cause of handling damage was assigned to the analysed ¿coil delivery wire kinked/bent as the defect appears to be associated with handling of the product or portion of the product during the procedure.
 
Event or Problem Description
It was reported that during procedure of venous sinus emboli of dural-avf, the coil (subject device) was advanced into the microcatheter, but there was strong resistance about 10 cm before the tip of the catheter, so it was retrieved.While retrieving the coil (subject device) , it got prematurely detached at the rhv(rotating hemostasis valve).The detached coil (subject device) was safely removed from the patient¿s anatomy by removing the rhv.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1.5MM X 4CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15795255
Report Number3008881809-2022-00569
Device Sequence Number3010266
Product Code HCG
UDI-Device Identifier04546540697998
UDI-Public04546540697998
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2025
Device Model NumberM0035421540
Device Catalogue NumberM0035421540
Device Lot Number236839151
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PHENOM17 MICROCATHETER(NON-STRYKER)
Patient SexPrefer Not To Disclose
-
-