We checked the returned unit and confirmed that the ccd module cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module.In addition, we confirmed that the light guide fiber bundle (lcb) broken, the insertion flexible tube (ift) buckled, the light guide cable buckled, the objective lens unit cracked, the operation channel leak, and the distal body worn out; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|