• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) 5356 VERITY DR, AB 20 55 184; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) 5356 VERITY DR, AB 20 55 184; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5356
Medical Device Problem Code Pacemaker Found in Back-Up Mode (1440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented for a follow-up in clinic.Upon interrogation, it was noted that the pacemaker was found in back-up vvi mode due to threshold testing.No intervention was performed.The patient was in stable condition.
 
Event or Problem Description
New information noted that the pacemaker was explanted and replaced.The patient was stable throughout the procedure.
 
Additional Manufacturer Narrative
As-received device was in bvvi backup mode and attempts to reload the product code was not successful due to low battery voltage.Electrical and mechanical analysis was performed, and the analysis showed that the implant duration meets projected longevity based on nominal settings indicating normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5356 VERITY DR, AB 20 55 184
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15795683
Report Number2017865-2022-45390
Device Sequence Number3113577
Product Code DXY
UDI-Device Identifier05414734006835
UDI-Public05414734006835
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2008
Device Model Number5356
Device Catalogue Number5356
Device Lot Number0002206064
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer11/14/2022
02/23/2023
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date12/02/2022
02/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TENDRIL LEAD.; TENDRIL LEAD.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient EthnicityHispanic
Patient RaceBlack Or African American
-
-