Model Number 211616
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an ultrasound guided breast biopsy through normal tissue, the device allegedly self activated.A coaxial was used.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 02/2025).
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one monopty disposable core biopsy instrument was returned for evaluation.On visual evaluation, the device appeared to be clean and the device was received in a condition where the arrow was in the ready window(i.E second rotational position).On functional evaluation, it was noted that the device did not self-activate.And it was noted that the device was able to be fully primed with no issues and the arrow was visible in the ready window.On functional testing, the device was cocked and fired into the chicken breast for 20 times with a total of 20 tests.All 20 samples were obtained with no issues.Therefore, the investigation for the reported self-activation is unconfirmed as the device was able to prime and fire properly with no issues.A definitive root cause for the alleged self-activation could not be determined based upon the provided information.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: d4 (expiry date: 02/2025), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an ultrasound guided breast biopsy through normal tissue, the device allegedly self activated.A coaxial was used.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOPTY
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15795921
Report Number2020394-2022-00896
Device Sequence Number11930767
Product Code KNW
UDI-Device Identifier00801741084720
UDI-Public(01)00801741084720
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number211616
Device Catalogue Number211616
Device Lot NumberREGQ1177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer01/02/2023
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-