• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INS SIZE4 13MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INS SIZE4 13MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5531-G-413
Medical Device Problem Codes Degraded (1153); Fracture (1260); Unstable (1667)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 10/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Knee revision due to instability.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Knee revision due to instability.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture, wear, and instability involving a triathlon insert was reported.The event was confirmed via evaluation of the returned device and clinician review of provided medical records.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that the insert is worn.The damage present on the posterior portion was consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.Clinician review: a review of the provided medical information by a clinical consultant indicated: "the office note states that the patient had significant instability in a previously replaced knee.The instability is again noted in the operative report.In addition, the surgeon states that there was gross wear and fracture on the "dorsomedial" surface of the tibial liner.He went on to remove the failed insert and replaced it with a thicker one.Revision total knee surgery for tka instability is confirmed.The root cause of this tka instability and tibial component wear cannot be determined by the information provided." product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to instability and fracture/wear of the triathlon insert.The fracture and wear of the insert were confirmed via evaluation of the returned device.A review of the provided medical records by a clinical consultant confirmed joint instability but could not identify root cause.Further information such as additional pre- and post-operative x-rays as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.It was reported that the patient had a "slip accident" and that the knee became instable after this incident.It is likely that this patient trauma contributed to the device failure.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INS SIZE4 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick, NJ NA
EI   NA
2018315000
MDR Report Key15795958
Report Number0002249697-2022-01625
Device Sequence Number14641164
Product Code MBH
UDI-Device Identifier07613327045819
UDI-Public07613327045819
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model Number5531-G-413
Device Catalogue Number5531G413
Device Lot NumberLDR669
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2022
Supplement Date Received by Manufacturer01/14/2023
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date02/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight98 KG
-
-