Catalog Number 03P76-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Hypoxia (1918); Respiratory Failure (2484)
Date of Event 10/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded a suspected descrepant hemglobin results on a 50 year old male patient with acute respiratory failure & hypoxia.There was no additional patient information available at the time of this report.Return product is not available for investigation.I-stat 30-oct-2022 22:00 22:03 29 mmhg 94 mmhg, lab 30-oct-2022 22:27 22:37 48.4 mmhg 184.50 mmhg.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 13-dec-2022.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for eg7+ cartridge lot n22165.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EG7+ CARTRIDGE
Common Device Name
EG7+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15796374
Report Number2245578-2022-00159
Device Sequence Number17341011
Product Code CHL
UDI-Device Identifier10054749000040
UDI-Public10054749000040
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K940918
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2023
Device Catalogue Number03P76-25
Device Lot NumberN22165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer12/13/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
-
-