MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE Back to Search Results
Catalog Number 03P76-25
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Hypoxia (1918); Respiratory Failure (2484)
Event Date 10/30/2022
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded a suspected descrepant hemglobin results on a 50 year old male patient with acute respiratory failure & hypoxia.There was no additional patient information available at the time of this report.Return product is not available for investigation.I-stat 30-oct-2022 22:00 22:03 29 mmhg 94 mmhg, lab 30-oct-2022 22:27 22:37 48.4 mmhg 184.50 mmhg.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 13-dec-2022.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for eg7+ cartridge lot n22165.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EG7+ CARTRIDGE
Type of Device
EG7+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15796374
MDR Text Key308014328
Report Number2245578-2022-00159
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K940918
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/11/2023
Device Catalogue Number03P76-25
Device Lot NumberN22165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
-
-