• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH PERIARTIC AIM ARM 4.5 LCP PROX TIB PL-RT; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH PERIARTIC AIM ARM 4.5 LCP PROX TIB PL-RT; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 03.120.001
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, while performing a proximal tibia, the surgeon felt there was issues with the aiming arm because the threaded towers were not able to reach the plate.Once the surgeon pushed harder, the towers were able to engage into the plate.It appears that when the plate was pulled to the bone with a screws, it pulled the plate away from the handle and further away from the tower.The surgery was successfully completed.Sales representative spoke with the surgeon after the case and said this handle replaced in case there was an issues with aiming arm.The handle was not lining up with the plate holes.It was checked prior to insertion and it lined up, however intra-op, after plate was brought to bone , it did not line up.No adverse events occurred.There was no surgical delay.The procedure was completed successfully.Concomitant device reported: unk - plates: trauma (part# unknown; lot# unknown; quantity: unknown) and unk - screws: trauma (part# unknown; lot# unknown; quantity: unknown).This report is for one (1) periartic aim arm 4.5 lcp prox tib pl-rt.This is report 2 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Mfr site postal code - (b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: investigation summary the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects on the surface of the periartic aim arm 4.5 lcp prox tib pl-rt, p/n: 03.120.001.A dimensional inspection was unable to be performed due to not being applicable to the complaint.A functional test to assemble the periart insrt hndl 4.5 lcp prx tib pl-rt, p/n: 03.120.002, to the returned periartic aim arm 4.5 lcp prox tib pl-rt, 03.120.001, was conducted.The coupling of the insertion handle was inserted in the aiming arm.After that, the coupling bolt was tightened.The devices assembled correctly.A functional test to assess the assembly to the plate was not conducted since the plate was not returned for evaluation.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the periartic aim arm 4.5 lcp prox tib pl-rt, p/n: 03.120.001 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a.Device history lot part: 03.120.001 lot no : 3113483 release to warehouse date: 24 april 2009 manufacturing site: werk hagendorf a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIARTIC AIM ARM 4.5 LCP PROX TIB PL-RT
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15796777
Report Number8030965-2022-09800
Device Sequence Number4546299
Product Code FZX
UDI-Device Identifier10886982073990
UDI-Public(01)10886982073990
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.120.001
Device Catalogue Number03.120.001
Device Lot Number3113483
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/24/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PERIART INSRT HNDL 4.5 LCP PRX TIB PL-RT; UNK - PLATES: TRAUMA; UNK - SCREWS: TRAUMA
Patient Age71 YR
Patient SexMale
-
-