EXACTECH, INC. EQUINOXE GLENOID, PEGGED ALPHA, MEDIUM; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
314-02-03 |
| Medical Device Problem Codes |
Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
10/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 5 years post op the initial right tsa, this 48 y/o male patient was revised due to a painful shoulder with loose pegged anatomic glenoid.Glenoid was revised to an equinoxe reverse.Patient was last known to be in stable condition following the event.Devices are not retuning, disposed of by hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation, concomitant device(s): equinox square torque define screw drive kit - 300-20-02, 4451186.Equinoxe replicator plate 1.5mm o/s - 300-10-15, 3627317.Equinoxe, humeral head tall, 44mm (alpha) - 310-02-44, 3695551.Humeral stem, 11mm - 300-01-11, 3835841.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the revision reported was likely the result of an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the glenoid component.Contributing factors to the reported loosening may include eccentric loading of the glenoid component and the patient¿s reported high activity level.However, this cannot be confirmed because the component was not returned for evaluation and radiographs were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6, h11 no device was returned for evaluation; the reported event was confirmed by review of photographs of the device.The review identified that properly functioning implants depend on their appropriate fixation to the bone; fixation is usually achieved by cementing the implant onto the bone.Although implants are firmly fixed at the initial shoulder replacement surgery, they may become loose over time in a multfactioral manner.This case appears to be a classic example of the ¿rocking horse¿ effect, which was first described in the literature by franklin et al.1 35 years ago.There is increased wear on the superior/posterior aspect of the glenoid component as a result of eccentric loading; this pattern is atypical and means that the humeral head was continuously contacting a small area of the glenoid implant, rather than properly translating across the glenoid face.This loading scenario was most likely caused by humeral head migration due to disruption of the concavity-compression mechanism secondary to an insufficient/failing rotator cuff, although other factors may have included incomplete glenoid seating, implant malposition, and/or the high activity level of the patient (as described in the hospital report).Over time, this eccentric loading on the rim of the glenoid generates a torque on the fixation surface that induces a stress at the bone-cement-implant interface, which can lead to interfacial failure and early glenoid loosening.Additionally, friction caused by the joint surfaces rubbing against each other wears away the surfaces of the implant, creating tiny particles that accumulate around the joint.In a process called aseptic (non-infected) loosening, the bond of the implant to the bone is also digested (a condition called osteolysis), which can weaken or even fracture the bone and exacerbate the effects of imbalanced joint loading on implant fixation.When the prosthesis becomes loose, the patient may experience pain, change in alignment, or instability.In the provided image, there is cement adhered to the pegs of the glenoid component.This appears consistent with implantation and removal.Operative notes and/or medical records were not provided for review of usage/ technique.A review of manufacturing data was performed and no discrepancies were found.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the glenoid component.Contributing factors to the reported loosening may include eccentric loading of the glenoid component and the patient¿s reported high activity level.However, this cannot be confirmed because the component was not returned for evaluation and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|