Model Number 989605488741 |
Device Problems
Mechanical Jam (2983); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
|
Event Description
|
A customer reported an x8-2t model transducer is not functional, it advance the angle then pops backwards.The field service engineer confirmed an articulation issue.There was no adverse patient or user impact associated with this event.
|
|
Manufacturer Narrative
|
A though evaluation of the transducer could not confirm the articulation issue as described by the customer.The transducer passed all performance testing, however, several cosmetic checks failed for the cable, scratches on the tee sensor window, and scratches on the mid-handle.Although there was evidence of cosmetic damage, the investigation confirmed the transducer was operating as designed with no malfunctions.
|
|