• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC EPIQ CVX; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC EPIQ CVX; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number 989605488741
Device Problems Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2022
Event Type  malfunction  
Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event Description
A customer reported an x8-2t model transducer is not functional, it advance the angle then pops backwards.The field service engineer confirmed an articulation issue.There was no adverse patient or user impact associated with this event.
 
Manufacturer Narrative
A though evaluation of the transducer could not confirm the articulation issue as described by the customer.The transducer passed all performance testing, however, several cosmetic checks failed for the cable, scratches on the tee sensor window, and scratches on the mid-handle.Although there was evidence of cosmetic damage, the investigation confirmed the transducer was operating as designed with no malfunctions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ CVX
Type of Device
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
thuy nguyen
22100 bothell everett highway
bothell, WA 98021
MDR Report Key15796843
MDR Text Key307597043
Report Number3019216-2022-00100
Device Sequence Number1
Product Code IYN
UDI-Device Identifier00884838088658
UDI-Public00884838088658
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132304
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number989605488741
Device Catalogue Number795231
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2022
Initial Date FDA Received11/15/2022
Supplement Dates Manufacturer Received05/15/2023
Supplement Dates FDA Received05/17/2023
Date Device Manufactured05/16/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-