• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH EPUMP FEED AND 1000MLFLUSH SET; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH EPUMP FEED AND 1000MLFLUSH SET; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 773662
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that during nutrition administration, air was evidenced in the device.The device was primed several times but the failure continued.Finally, the bag was swapped.There was no patient injury, medical intervention, or adverse reaction associated with this event.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on (b)(6), 2021.There were no physical samples received for investigation, but a video was provided.The video was reviewed, and air bubbles were observed through the entire line.The reported condition was confirmed.In addition, both the feed and flush bags were observed to be completely empty.The investigation was carried out with the multifunctional team, all processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.Without a physical sample to evaluate, a root cause could not be determined.If the sample is received at a later date, the investigation will be updated to reflect our findings.A corrective action is not applicable at this time.Functional testing and visual inspections are performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPUMP FEED AND 1000MLFLUSH SET
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15796904
Report Number9612030-2022-03464
Device Sequence Number4790992
Product Code LZH
UDI-Device Identifier30884521022106
UDI-Public30884521022106
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number773662
Device Catalogue Number773662
Device Lot Number2120233564
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-