MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT CR CEM FEM CR CEM LEFT SZ 4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT CR CEM FEM CR CEM LEFT SZ 4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number TRULIANT CR CEM FEM CR CEM LEFT SZ 4
Medical Device Problem Codes Noise, Audible (3273); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 1.5 years post initial left side tka, the 63 y/o female patient complained of a clicking sound in her knee, the femur showed radiolucent lines on films.Surgeon revised the loose femur to a size 4 ps femur.There was no breakage of a device.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain x-rays or images.The devices are not available for evaluation due to hospital policy will not allow for the return of the explant.No additional patient information provided.
 
Additional Manufacturer Narrative
Concomitant medical products: truliant tib imp crc insert sz 4, 9mm.Cat# 02-022-51-4009, serial# (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
(h3) the revision reported was likely the result of an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision radiographs were unable to be obtained.The most probable root cause associated with the reported event of ¿loosening - femoral¿ is associated with weakened integration of the femoral component at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6, h11 the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision radiographs were unable to be obtained.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT CR CEM FEM CR CEM LEFT SZ 4
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15797022
Report Number1038671-2022-01472
Device Sequence Number4674261
Product Code JWH
UDI-Device Identifier10885862320643
UDI-Public10885862320643
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K170240
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRULIANT CR CEM FEM CR CEM LEFT SZ 4
Device Catalogue Number02-020-13-0240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer03/27/2023
04/15/2026
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date04/17/2023
05/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADVANCED PATELLA 35MM 3 PEG IMPLANT.; FAN SAWBLADE EZ2213F.M63 90X22 1.27MM STRKR 5/6/7.; TRULIANT TIB FIT TRAY CEM SZ 4F / 4T.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight56 KG
-
-