| Model Number |
IPN917233 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
The report states that "after 10 minutes of use, a rupture of the catheter body outside the patient during angiographic injection with the usual volume and flow occurred.As a result, another catheter was inserted in the vein jugular left".Additional information states that the second catheter was inserted at the same insertion site.No medical intervention required.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The lot number reported is 16f21l0081.The lot number of the returned device is suspected to be from the reported lot; however, no original packaging/label was returned.Returned for investigation was a 5fr.50cm berman catheter without the original packaging.The sample was returned in a cardboard box and was in a sealed biohazard bag.Upon return, the catheter body was immediately noted ruptured near the junction hub; the body was noted ruptured from approximately 59.2cm to 59.9cm from the distal tip of the catheter.The inflation lumen stopcock was in the open position.The recommended volume capacity of the balloon is 0.75cc.The supplied control stroke syringe was not returned with the sample.Upon microscopic inspection, a wrinkle was noted on the balloon; no damage was noted to the balloon.No condensation was noted in the inflation lumen.Dried blood was noted on the exterior of the sample and within th e berman holes (injection lumen) near the catheter tip.No other damage or abnormalities were noted.The inflation lumen was injected with 0.75cc of air using a lab inventory control stroke syringe.The balloon inflated symmetrically.One side of the balloon measured approximately 4mm.The other side measured approximately 4mm.The balloon did meet specifications of radius ratio less than or equal to 2.0.The catheter's injection lumen was aspirated/flushed and air was noted leaking from the damaged/ruptured catheter body.The inflation lumen was injected with 0.75cc of air using a lab inventory control stroke syringe.The balloon inflated symmetrically.The balloon deflated in less than 3 seconds when the syringe was removed.No pull away was noted after the tug test.The balloon was placed in water and air was injected into the inflation lumen again.No leak was noted.The damaged/ruptured catheter body did not result in damage to the inflation lumen.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for "rupture of the catheter body" is confirmed.The catheter body was found ruptured near the junction hub upon receipt of the sample.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the ruptured catheter body.The root cause of the ruptured catheter body could not be confidently determined, but a potential cause is high pressure from injection.The complaint finding is considered isolated.No further action required at this time.This will be monitored for any developing trends.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
The report states that "after 10 minutes of use, a rupture of the catheter body outside the patient during angiographic injection with the usual volume and flow occurred.As a result, another catheter was inserted in the vein jugular left".Additional information states that the second catheter was inserted at the same insertion site.No medical intervention required.Patient condition reported as "fine".
|
| |
|