| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
10/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Measure: this issue only impacts this device and does not impact any other fielded devices at this time.Analyze: cas, (b)(6) reviewed case #875.Centration placement is good with adequate suction to the eye.Minor patient movement was noted throughout.Frame #25 showed a 32 degree ak starting at an axis of 358 degrees.Ak was completed in frame # 33.Root cause: patient movement likely contributed to the poor localized imaging.Resulting in the corneal perforation.Laser functioned as design.
|
| |
|
Event or Problem Description
|
|
Issue: on (b)(6) 2022, ((b)(4)) reported to (b)(6) that "dr.(b)(6) feels like his arcuate incision may have perforated and he would like the case reviewed." dr.(b)(6) sutured the ak and proceeded with no complications.
|
| |
|