• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDERITE SINGLE-USE STYLET - SMALL; STYLET, TRACHEAL TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDERITE SINGLE-USE STYLET - SMALL; STYLET, TRACHEAL TUBE Back to Search Results
Model Number 0803-0118
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Unintended Extubation (4564)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
The following was reported by a user facility via medsun mandatory and voluntary report number (b)(4) which stated that during an emergency care procedure of a 9-month old patient, using a gliderite single-use stylet - small, the stylet snapped prompting a repeat intubation attempt with a different tube and stylet.The patient was bagged between intubation attempts with no difficulty, was successfully intubated, and transferred to the pediatric intensive care unit (picu).No harm to the patient was reported.
 
Additional Manufacturer Narrative
The initial reported device used during the patient procedure was not retained by the facility and therefore will not be returning to verathon for evaluation.However, the facility plans on returning additional samples from their stock for evaluation.Initial follow-up information received from the facility reported that the stylet broke at the bended section while trying to adjust it during intubation.The broken piece was retrieved but they did not indicate from where.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
 
Additional Manufacturer Narrative
Replacement gliderite single-use stylets - small were provided to the customer and eight (8) unopened stylets from the same lot of the reported stylet were returned to verathon.The device used during the reported event was not retained by the facility and therefore was not able to be returned to verathon for evaluation.Since the reported device was not returned to verathon for evaluation, the cause could not be determined.However, based on follow-up information received from the facility, it is likely that the cause of the reported event may have been due to the stylet being bent while trying to adjust it during intubation.The gliderite single-use stylets are not to be bent or reshaped, as the shape of the stylet is designed to complement the curve of glidescope video laryngoscopes.The model and size of the et tube used with the stylet during the reported event is unknown.The gliderite single-use stylet (small) instructions for use states: "the stylet is for use in et tubes between 3.0 and 4.0 mm in internal diameter.Note: there is no guarantee that instruments selected solely using these intrument deminsions will be compatible in combination." review of complaint history for the reported gliderite single-use stylet - small, lot number "gs77314," identified one (1) other complaint reported to verathon in august 2021.Trending analysis for the gliderite single-use stylet - small does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized by verathon.Corrective action is currently not required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDERITE SINGLE-USE STYLET - SMALL
Common Device Name
STYLET, TRACHEAL TUBE
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key15798141
Report Number9615393-2022-00209
Device Sequence Number12177646
Product Code BSR
UDI-Device Identifier10879123005967
UDI-Public011087912300596710GS7731417240331
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number0803-0118
Device Catalogue Number0270-0916
Device Lot NumberGS77314
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 MO
Patient SexFemale
Patient Weight6 KG
Patient RaceWhite
-
-