| Model Number |
C501 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a questionable low valproic acid (valp2) result from the cobas 6000 c501 module.The initial result was 0.8 ug/ml with a data flag and was reported outside of the laboratory.The repeat results were 30.1 ug/ml and 45.4 ug/ml.The repeat results on 27-oct-2022 were 51.1 ug/ml and 45.2 ug/ml.The reagent lot number and expiration date were requested but were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.The calibration and qc data were acceptable.The analyzer alarm trace did not show any serious alarms.
|
| |
|
Additional Manufacturer Narrative
|
|
Medwatch fields a2 - age at time of event, 2 - date of birth, a3 - sex, b3 - date of event, and medications were updated.The results of 30.1 ug/ml and 45.4 ug/ml were obtained on (b)(6) 2022.The reagent lot number was 598405.
|
| |
|