| Model Number |
110-150 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported right side recall and rupture.Further reported was "intrauterine carcinoma of breast" which has been deemed not device related.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken in anterior side assessed as surgical damage and other pattern broken in posterior side assessed as unidentified (tear) opening (shell thickness within specification).Anxiety¿product/procedure: unable to observe since it is a medical event and is not related to the device.Cancer brest-ndr: not applicable as the event is not related to the device.Additional observations: creases are observed.Wear abrasion is observed.Missing shell assessed as inconclusive.
|
| |
|
Event or Problem Description
|
|
Patient reported right side recall and rupture.Further reported was "intrauterine carcinoma of breast" which has been deemed not device related.Device has been explanted and replaced.
|
| |
|