Model Number 434899
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Cardiac Arrest (1762); Cardiac Tamponade (2226)
Date of Event 10/13/2022
Type of Reportable Event Death
Event or Problem Description
After placing a stent in the mid lad in a pre-dilated total occlusion and subsequently performed post dilatation, a type iii perforation was detected.The pk papyrus covered stent system was selected for treatment of the perforation.The perforation worsened after balloon dilation and pk papyrus deployment.The tear was extended in both directions, proximally and predominantly distal to the vessel.Cardiac tamponade and pea arrest occurred.Attempted ecmo placement, without success.Patient expired in the cath lab.
 
Additional Manufacturer Narrative
(b)(6) 2022 opened in error.Wrong device reported.Closing report and opening new report on correct device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 3.0/26
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15798387
Report Number1028232-2022-05817
Device Sequence Number4754150
Product Code NIV
UDI-Device Identifier07640130447271
UDI-Public(01)07640130447271(17)2309
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number434899
Device Catalogue NumberSEE MODEL NO.
Device Lot Number08203280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age66 YR