MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 0580-1-443
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 10/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
The following was reported: hip revision surgery due to patient's exeter stem 44 n3 being broken at neck level.
 
Event or Problem Description
The following was reported: hip revision surgery due to patient's exeter stem 44 n3 being broken at neck level.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a exeter stem was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event could not be confirmed as insufficient information was provided.Further information such as device return, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
The following was reported: hip revision surgery, due to patient's exeter stem 44 n3, being broken at neck level.
 
Additional Manufacturer Narrative
Reported event: an event, regarding crack/fracture involving a exeter stem was reported.The event was confirmed, via evaluation of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed, as the device was not returned.Clinician review: a review of the provided medical records, by a clinical consultant indicated.I can confirm, that the patient had the primary hip replacement, since i was able to review the operation report.I can also confirm, with secondary evidence that the patient had a revision of the left total hip arthroplasty, since i was able to review a radiology report, which stated there was an "interval revision of the left sided total hip arthroplasty.No x-ray documentation was provided.And no revision operation report was provided.Root cause analysis: the root cause of this event cannot be determined, with certainty.Causes of fracture of a prosthetic femoral neck are multifactorial including surgical technique, especially with a cemented implant regarding the cement mantle, patient factors including bmi and activity level.This patient apparently was quite active and had a bmi of over 34.Also to be considered, are implant factors.Product history review: review of the device history records, indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event was confirmed.Via review of the provided medical records.Which indicated, "i can confirm, that the patient had the primary hip replacement, since i was able to review the operation report.I can also confirm, with secondary evidence that the patient had a revision of the left total hip arthroplasty, since i was able to review a radiology report.Which stated, there was an "interval revision of the left sided total hip arthroplasty".No x-ray documentation was provided.And no revision operation report was provided.Root cause analysis: the root cause of this event cannot be determined, with certainty.Causes of fracture of a prosthetic femoral neck are multifactorial including surgical technique.Especially, with a cemented implant regarding the cement mantle, patient factors including bmi and activity level.This patient apparently, was quite active and had a bmi of over 34.Also to be considered, are implant factors".No further investigation for this event is possible, at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a exeter stem was reported.The event was confirmed via evaluation of the provided medical records.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of the provided medical records by a clinical consultant indicated: i can confirm that the patient had the primary hip replacement since i was able to review the operation report.Originally, i was able to confirm with secondary evidence that the patient had a revision of the left total hip arthroplasty since i was able to review a radiology report which stated there was an "interval revision of the left sided total hip arthroplasty." now, with the additional documents supplied for the addendum i can confirm that the patient underwent revision of the left total hip arthroplasty since i was able to see x-ray with the revision implant in place.Root cause analysis: the root cause of this event cannot be determined with certainty.Causes of fracture of a prosthetic femoral neck are multifactorial including surgical technique, especially with a cemented implant regarding the cement mantle, patient factors including bmi and activity level.This patient apparently was quite active and had a bmi of over 34.Also to be considered are implant factors.The additional information does not change the root cause analysis.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event was confirmed via review of the provided medical records which indicated, " i can confirm that the patient had the primary hip replacement since i was able to review the operation report.Originally, i was able to confirm with secondary evidence that the patient had a revision of the left total hip arthroplasty since i was able to review a radiology report which stated there was an "interval revision of the left sided total hip arthroplasty." now, with the additional documents supplied for the addendum i can confirm that the patient underwent revision of the left total hip arthroplasty since i was able to see x-ray with the revision implant in place.Root cause analysis: the root cause of this event cannot be determined with certainty.Causes of fracture of a prosthetic femoral neck are multifactorial including surgical technique, especially with a cemented implant regarding the cement mantle, patient factors including bmi and activity level.This patient apparently was quite active and had a bmi of over 34.Also to be considered are implant factors." no further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER V40 STEM 44MM NO 3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key15798863
Report Number0002249697-2022-01633
Device Sequence Number19646394
Product Code JDI
UDI-Device Identifier04546540153326
UDI-Public04546540153326
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/25/2025
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number0580-1-443
Device Catalogue Number0580-1-443
Device Lot NumberG7279115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/19/2022
Supplement Date Received by Manufacturer03/30/2023
03/21/2024
05/29/2025
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date04/24/2023
04/11/2024
06/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexMale
-
-