(b)(4).Date of event: only event year known: 2022.Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Additional information was requested, and the following was received: did the device lock-out (no staples deployed and no cut line started)? unknown was there any difficulty opening the device? yes.It was difficulty opening during the surgery.If yes, how was the device removed from the patient? use another gun to cut around the tissue.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
|
(b)(4), date sent: 12/15/2022, d4: batch # 569a88.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the one sc45a device was returned with joint cover damaged, the closing trigger and anvil in the closed position.Also, the firing mechanism in the return stroke with the knife not fully back.One gst45d reload was loaded in the device.The cartridge reload was received fully fired and with damage on cartridge deck.In addition, a metal wire was found lodged behind the knife, resulting in the firing mechanisms jamming.The knife was returned to home position, the metal wire was removed and the device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.The event reported was confirmed and it is related to improper use of the device.To mitigate the potential for staples getting into the cartridge and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and cartridge jaw in sterile solution and then wipe the anvil and cartridge jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.The damage to the reload is consistent with the device being clamped over a hard object.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.The damaged on the joint cover, it is possible that the device was attended to be pulled out from a trocar in the articulated position, resulting in the edge of the trocar damaging the joint cover.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 569a88, and no non-conformances were identified.
|