| Model Number |
MSERIES CCT |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
Necrosis (1971)
|
| Date of Event |
10/25/2022
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a 45-year-old male patient, the device failed to discharge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired; however, they do not believe this was device associated.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, defib testing, a hot/cold environmental testing, and verification that the device would automatically disarm after one minute without duplicating the report.The device was returned to the customer without going through recertification as a non-pma apprvoed device and the customer was informed that this device is retired and no longer on the fda-approved aed list.No trend is associated with reports of this type.
|
| |
|