• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD INSYTE¿ IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD INSYTE¿ IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381212
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the tip of the bd insyte¿ iv catheter broke when advancing it in the patient's vein.This occurred with 3 catheters.The following information was provided by the initial reporter: "it was reported that before they place a catheter they advance the stylette all the way up the catheter and this particular lot have been consistently breaking at the tip when they do this.".
 
Event or Problem Description
It was reported that the tip of the bd insyte¿ iv catheter broke when advancing it in the patient's vein.This occurred with 3 catheters.The following information was provided by the initial reporter: "it was reported that before they place a catheter they advance the stylette all the way up the catheter and this particular lot have been consistently breaking at the tip when they do this.".
 
Additional Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 11-nov-2022.H.6.Investigation summary: our quality engineer inspected the 1 photo and 7 representative samples submitted for evaluation.The reported issue of catheter tip integrity was not confirmed upon inspection of the samples.Analysis of the sample showed that they were all within manufacturing specifications and no damages were observed.Bd could not determine a manufacturing root cause since the defect was not confirmed.It is recommended to send the sample that was involved in the reported incident.Production records were reviewed, and this batch was in compliance with our product specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15799660
Report Number8041187-2022-00663
Device Sequence Number19585794
Product Code FOZ
UDI-Device Identifier00382903812127
UDI-Public(01)00382903812127
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2025
Device Catalogue Number381212
Device Lot Number0169262
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-