Catalog Number 309572
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter phone#: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that an unspecified number of bd luer-lok¿ syringes with the bd precisionglide¿ needle had issues connecting to the syringe during use and leaked medication as a result.The following information was provided by the initial reporter: "others would not seal correctly where needle is threaded to syringe causing difficulty drawing meds and losing meds when administering (so not knowing if proper dose was received).".
 
Event or Problem Description
It was reported that an unspecified number of bd luer-lok¿ syringes with the bd precisionglide¿ needle had issues connecting to the syringe during use and leaked medication as a result.The following information was provided by the initial reporter: "others would not seal correctly where needle is threaded to syringe causing difficulty drawing meds and losing meds when administering (so not knowing if proper dose was received).".
 
Additional Manufacturer Narrative
A device history review was conducted for lot number 2048232.A review showed no rejected inspections or quality issues during the production that could have contributed to the reported defect.One photo was provided to our quality team for investigation.Based on the investigation and with no physical sample analysis the symptom reported by the customer could not be confirmed.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ SYRINGE WITH THE BD PRECISIONGLIDE¿ NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15800081
Report Number1213809-2022-00991
Device Sequence Number3010523
Product Code FMF
UDI-Device Identifier00382903095728
UDI-Public(01)00382903095728
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K110771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number309572
Device Lot Number2048232
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-