| Model Number |
NT4F19115 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device was not sent to olympus.A review of the device history record found no deviations that could have caused or contributed to the broken device.The device shipped according to specifications.As part of the in-process testing, the minimum tensile strength of the basket and the connection between the basket and the tensile wire is tested to 100%.Based on the results of the legal manufacturer's investigation, without physical device evaluation, a definitive root cause of the reported issue could not be identified.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
A query of the manufacturer and user facility device experience (maude) database found a voluntary report that states the subject device broke upon normal ureteral stone retrieval.According to the report, all equipment pieces were intact and no harm was done to the patient.Because the voluntary report does not include contact information for the submitter, additional information cannot be requested.
|
| |
|
Search Alerts/Recalls
|