| Model Number |
625090H |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This is 1 of 2 reports for the handle used during this event and is linked to manufacturer report# 3014334038-2022-00257: a facility reported that during a laparoscopic cholecystectomy, the trial detach roto-cam handle/shaft (625090h) did not function properly; the insert did not fit/link with the handle.No patient injury, death or surgical delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The loaner detach roto-cam handle/shaft (625090h) has been returned for evaluation.The device history record (dhr) was reviewed, and no anomalies related to the reported failure were observed.Visual evaluation noted that the roto-cam handle/shaft was received in used condition with the shaft unable to link to the handle due to wear/environmental damage causing misalignment.The reported complaint was confirmed.No manufacturing, workmanship or material deficiency has been identified.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Search Alerts/Recalls
|