MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-65-080-120-P6
Medical Device Problem Codes Difficult to Remove (1528); Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified, mildly tortuous, 90% stenosed, de novo lesion in the superficial femoral artery (sfa).The 6.5x80mm supera peripheral self-expanding stent system (sess) was advanced to the lesion.Upon deployment, the thumbslide stopped and it was noted that the tip of the delivery system was caught on the stent.The stent was only partially deployed.It was decided to remove the device including the stent.Resistance was noted so the stent delivery system was removed with the sheath.It is unknown if the stent was removed or remains in the anatomy.The procedure was completed with a drug coated balloon.No additional information was provided.
 
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
Subsequent to the initial report, the following information was received: the stent was only partially deployed.It was decided to remove the device including the stent.Resistance was noted so the stent delivery system was removed with the sheath.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, it is likely that interactions with the heavily calcified, mildly tortuous and 90% stenosed anatomy resulted in the ratchet being unable to properly engage and release the stent from the delivery system; thus resulting in reported activation/ deployment failure and the reported mechanical jam.During removal interaction with the heavily calcified, mildly tortuous and 90% stenosed anatomy resulted in the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15802055
Report Number2024168-2022-11630
Device Sequence Number7074588
Product Code NIP
UDI-Device Identifier08975370026368
UDI-Public08717648226267
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberS-65-080-120-P6
Device Catalogue NumberS-65-080-120-P6
Device Lot Number2031061
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-