ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
TRM-405 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.Device evaluation: analysis of the returned device identified: weight to the spec.Bubbles in the gel observed after autoclave cycle the unit is not rupture.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Patient representative reported right side concern with product.Later, healthcare professional reported, an extracapsular rupture and exchange from a textured to smooth breast implant due to the patient¿s concern with the product.The device as been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|
|
|