• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. GRASPING FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. GRASPING FORCEPS Back to Search Results
Model Number FG-46U-1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 10/02/2009
Type of Reportable Event Serious Injury
Event or Problem Description
Olympus reviewed the following literature titled "usefulness of a second endoscopic arm to improve therapeutic endoscopy in the lower gastrointestinal tract.Preliminary experience ¿ a case series".The aim of our study was to test the feasibility of utilizing a second endoscopic arm to exert counter-traction to lift the lesion away from the cutting plane, ameliorating submucosal exposure.Total eight patients were included in this study.Three complications occurred and were managed endoscopically.One patient experienced a small perforation in the sigmoid colon.Two patients presented with delayed bleeding.Applying counter-traction with a second endoscopic arm can facilitate submucosal dissection of distal colorectal lesions.Type of adverse events/number of patients, perforation - 1 patient, delayed bleeding - 2 patients.This literature article requires 9 reports.The related patient identifiers are as follows: (b)(6) for fg-48l-1 pt #4: perforation, (b)(6) for cf-q165i pt #4: perforation, (b)(6) for fg-46u-1 pt #4: perforation, (b)(6) for fg-48l-1 pt #5: delayed bleeding, (b)(6) for cf-q165i pt #5: delayed bleeding, (b)(6) for fg-46u-1 pt #5: delayed bleeding, (b)(6) for fg-48l-1 pt.#7: delayed bleeding, (b)(6) for cf-q165i pt.#7: delayed bleeding, (b)(6) for fg-46u-1 pt.#7: delayed bleeding.This medwatch report is for patient identifier (b)(6).There is no report of any olympus device malfunction in any procedure described in this study.
 
Additional Manufacturer Narrative
The authors report gf um160 was used to detect infiltration of the muscularis propria and/or the presence of perivisceral lymph nodes at preliminary endoscopic ultrasound which contraindicated endoscopic treatment and therefore excluded from the study.Xst-28ch-m is prototype device not cleared for market.The suspect device has not been returned to olympus for evaluation.The investigation is in process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The following additional information was received from the physician: ¿the adverse events described in the article are attributable neither to the olympus instrumentation nor to its malfunctioning, but rather to the method of submucosal dissection itself which, as is well known from the literature, is an invasive technique and subject to complications such as those described in the article.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRASPING FORCEPS
Common Device Name
GRASPING FORCEPS
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15802435
Report Number9614641-2022-00628
Device Sequence Number12165588
Product Code FFL
UDI-Device Identifier04953170031724
UDI-Public04953170031724
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K962387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFG-46U-1
Device Lot NumberUNKNOWN(LITERATURE)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/15/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NON-OLYMPUS, ROTH NET RETRIEVAL NET, SN UNKNOWN; OLYMPUS, CF-Q165I, SN UNKNOWN; OLYMPUS, FG-48L-1, LOT UNKNOWN
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age71 YR
Patient SexFemale
-
-