| Model Number |
C503 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Precision testing was performed with acceptable results.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with gluc3 (glucose hk gen.3) on a cobas c 503 analytical unit.The sample initially resulted in a gluc3 value of 0.6 mmol/l and repeated as 5.6 mmol/l.The gluc3 reagent lot was 62948701 with an expiration date of 31-jul-2023.It is not known which result is deemed correct.The questionable result was not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|