BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7133
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the shaft break occurred.A 2.00mm x 15mm emerge balloon catheter was selected for use.However, during preparation, the catheter broke before going into the body.The physician pulled a new device, and the procedure was completed.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Date of event- used the first date of the month of the aware date as no event date was provided.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of the emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.The balloon was tightly folded.There were numerous kinks throughout the hypotube of the device.There was a complete hypotube separation 82cm distal of the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Product analysis confirmed the reported event, as the hypotube was separated.B3.Date of event- used the first date of the month of the aware date as no event date was provided.
 
Event or Problem Description
It was reported that the shaft break occurred.A 2.00mm x 15mm emerge balloon catheter was selected for use.However, during preparation, the catheter broke before going into the body.The physician pulled a new device, and the procedure was completed.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15803971
Report Number2124215-2022-47383
Device Sequence Number5401934
Product Code LOX
UDI-Device Identifier08714729806196
UDI-Public08714729806196
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2025
Device Model Number7133
Device Catalogue Number7133
Device Lot Number0029376167
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-