• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL FLEXABILITY¿ ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FLEXABILITY¿ ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number A-FASE-DF
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the premature ventricular contraction procedure, electrodes 3-4 appeared to fold in 3d and a delay occurred.The systems were restarted which did not resolve the issue.Another ablation catheter was used to complete procedure with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
One quadripolar, bi-directional, curve d-f, flexability sensor enabled ablation catheter was received for evaluation.Electrodes 3-4 met specifications for acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported display issue and subsequent delay remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXABILITY¿ ABLATION CATHETER, SENSOR ENABLED¿
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15804765
Report Number3008452825-2022-00688
Device Sequence Number20209000
Product Code OAD
UDI-Device Identifier05415067020062
UDI-Public05415067020062
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberA-FASE-DF
Device Catalogue NumberA-FASE-DF
Device Lot Number8112813
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-