| Model Number |
SC941 MAXIMA 4 WHEEL |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
Provider alleges consumer was driving down a sidewalk when the weld on the seat post allegedly broke allegedly causing the consumer to fall out of the chair.
|
| |
|
Additional Manufacturer Narrative
|
|
It was found that on the lower seat post weldment (mec1904971) the weld penetration were insufficient see (ecr05606).
|
| |
|
Event or Problem Description
|
|
Provider alleges consumer was driving down a sidewalk when the weld on the seat post allegedly broke allegedly causing the consumer to fall out of the chair.
|
| |
|