• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL VASCULAR LIN STAP; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL VASCULAR LIN STAP; STAPLE, IMPLANTABLE Back to Search Results
Model Number TX30V
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing record evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure, when the surgeon was using the device, the cartridge was not fully fired in the process of firing.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 01/13/2023.D4: batch # u5deou.Investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample determined that the xr30v reload arrived with only 12 staples present and protruding.Also, drivers were noted partially advanced.Due to the condition of the reload no functional test was performed.In addition, the device was not received for analysis.It should be noted that when manipulating the device only the closing trigger should be grasp until ready to fire the device.Do not grasp the firing trigger before the device is to be fired.If the firing trigger is manipulated it could cause the staples to deploy partially or completely resulting in malformed staples and a premature lockout situation.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch u5deou, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15805507
Report Number3005075853-2022-08004
Device Sequence Number12165611
Product Code GDW
UDI-Device Identifier10705036003335
UDI-Public10705036003335
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model NumberTX30V
Device Catalogue NumberTX30V
Device Lot NumberU5DEOU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer12/26/2022
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
XR30V
Patient SexUnknown
-
-