ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TRANSPAC¿ IT MONITORING KIT, 27", 30 ML FLUSH DEVICE, 10 CC CONTAMINATION SYRING; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR Back to Search Results
Model Number 42800-39
Medical Device Problem Codes Disconnection (1171); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for evaluation; however, testing has not yet been completed.
 
Event or Problem Description
The event occurred on an unspecified date and involved a transpac¿ it monitoring kit, 27", 30 ml flush device, 10 cc contamination syringe and needleless valve.The reporter stated that a picu/nicu experienced a disconnect of the intravenous (iv) administration extension set at the bonding point on their transpac it transducer.There was no patient involvement or patient harm reported.
 
Additional Manufacturer Narrative
The reported complaint of separation was confirmed on the returned set.An image was provided by the customer showing the area of the defect.During visual inspection, the pvc tubing was found separated from the transducer luer pocket.The pvc tubing was observed to be tacky.The probable cause of the pvc tubing being tacky was due to the tubing not being fully cured during assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPAC¿ IT MONITORING KIT, 27", 30 ML FLUSH DEVICE, 10 CC CONTAMINATION SYRING
Common Device Name
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key15805717
Report Number9617594-2022-00302
Device Sequence Number5402011
Product Code DRS
UDI-Device Identifier00887709098169
UDI-Public(01)00887709098169(17)250901(10)8602165
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K052828
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2025
Device Model Number42800-39
Device Catalogue Number42800-39
Device Lot Number8602165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer11/30/2022
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-