MENTOR TEXAS MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number SDC120H
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a 375cc mentor artoura plus, smooth, high profile tissue expander being used in a breast surgery was found to have shell irregularities during the operation.No adverse event was experienced by the patient.The surgeon used backup devices, and there were no reported patient consequences or procedural delays.
 
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On december 29, 2022, mentor completed evaluation of the device.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that the artoura plus sm te hp 375cc tissue expander had an irregularity between the shell and patch on the anterior view.Leak testing was performed, according to the mentor procedure, and no leak sites were detected during the analysis.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through mentor¿s quality system.While this complaint was initiated for a defect that appeared to be a scratch, through the assessment it was identified to be an excess material as part of the manufacturing process, that will not cause injury and is unlikely to render the product unusable or difficult to use.There are inspections at various stages of the manufacturing process, however excess material is a normal variation that can occur.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b))(4).
 
Additional Manufacturer Narrative
On january 18, 2023, mentor received an updated serial number of the device.Relevant fields have been updated.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key15805907
Report Number1645337-2022-13881
Device Sequence Number5471620
Product Code LCJ
UDI-Device Identifier10081317028199
UDI-Public10081317028199
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K161176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2026
Device Model NumberSDC120H
Device Catalogue NumberSDC120H
Device Lot Number9672082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer12/29/2022
01/18/2023
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date12/30/2022
02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-