| Model Number |
4005472 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation into the reported event is in process; a follow-up mdr will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the pinch clamp for their endo smartcap tubing is not staying closed allowing water to leak onto the floor.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation determined that the cause of the tubing not staying clamped was attributed to the tip (point) of the clip.The tip is seated within the "teeth" of the clip to keep the clip engaged/closed.As the tip was not sharp/pointy enough, this allowed the point to detach from the teeth resulting in the reported issue.The issue appears to be isolated to a very small number of clips.We have notified our supplier, and will continue to monitor post-market data for trends.
|
| |
|