| Model Number |
DE2110X13B |
| Medical Device Problem Code |
Tidal Volume Fluctuations (1634)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator was not delivering the set volume.There was no harm or injury reported.The manufacturer is currently investigating this issue and a follow-up report will be filed when the investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
It was initially reported, the manufacturer received information alleging a ventilator was not delivering the set volume.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's machine flow sensor needs replaced due to contamination.
|
| |
|