|
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection of the returned device found that it is not in its original packaging.The insertion device, fractured anchor and suture materials were returned.The anchor is fractured in a spiral from the suture window down.There is debris on all returned items.Based on the condition of the product material found during visual inspection, additional material testing is not required.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.As of the date of this medical investigation, the requested clinical documentation has not been provided; no patient injury has been alleged.Therefore, no further clinical/medical assessment is warranted.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
|
It was reported that during a rotator cuff tear refixation, the 5.5mm twinfix anchor broke during deployment.The broken part stayed on the shaft.The procedure was completed with a non-significant surgical delay using a back-up device in the originally drilled bone hole.No further complications were reported.Results of investigation found biological debris on the insertion device, fractured anchor and suture materials.
|