MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3544251 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Granuloma (1876); Deformity/ Disfigurement (2360); Breast Discomfort/Pain (4504)
|
| Date of Event |
05/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast surgery with a 425cc mentor memorygel breast implant and experienced a rupture on the right side post-operatively.As a result, the patient underwent explantation, but a specific date has not been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
On november 23, 2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.An updated patient identifier, date of birth, and event date were received, as well as a confirmed date of explant, and additional information indicating the patient experienced a granuloma and breast pain post-operatively.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On december 14, 2022, mentor completed evaluation of the device.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the sx mpp gel 425cc breast implant had a tear on the anterior view, measuring approximately 1.1 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in an area of the tear, measuring approximately 0.1 cm.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.The cause of the rupture in the remaining area of the tear could not be identified.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.Granulomas are noncancerous nodules that are a common tissue reaction to the presence of a variety of foreign materials, such as silicone.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|