Model Number RLT311415
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 11/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2017, this patient underwent endovascular treatment for an abdominal aortic aneurysm and was implanted with gore® excluder® aaa endoprostheses featuring c3® delivery system the patient tolerated the procedure.On (b)(6) 2022, the patient underwent follow-up imaging and the physician reported that the aneurysm had grown proximally and now involved the renal arteries and the descending thoracic aorta.The physician plans to reintervene on (b)(6) 2022 to snorkel the renal arteries and extend coverage proximally into the descending thoracic aorta with gore® tag® thoracic stent grafts.The physician reported no endoleak just enlargement or elongation of the aneurysm.Additionally, it was unknown if there was migration of the device as a different physician performed the index procedure and it is unknown if placement of the original trunk component had been landed just distal to the lowest renal artery.
 
Additional Manufacturer Narrative
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H6: updated code.
 
Additional Manufacturer Narrative
H6: additional information.
 
Additional Manufacturer Narrative
H.6.Investigation conclusions: code d12 added: according to the gore® excluder® aaa endoprosthesis instructions for use (ifu),  additional considerations for patient selection include but are not limited to: aneurysm enlargement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15807877
Report Number3007284313-2022-02212
Device Sequence Number5623463
Product Code MIH
UDI-Device Identifier00733132619085
UDI-Public00733132619085
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/03/2019
Device Model NumberRLT311415
Device Catalogue NumberRLT311415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer11/10/2022
11/10/2022
03/03/2023
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date11/23/2022
01/12/2023
03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Other;
Patient Age70 YR
Patient SexMale
-
-